Recall Observatory FDA recall evidence

Device product

00597001303 Femur CR PRC FEM COMP SIZE CML Rx, Sterile; 00597001304 Femur CR PRC FEM COMP SIZE CMR Rx, Sterile; 00597001403 Femur CR PRC FEM COMP SIZE DML Rx, Sterile; 00597001404 Femur CR PRC FEM COMP SIZE DMR Rx, Sterile; 00597001503 Femur CR PRC FEM COMP SIZE EML Rx, Sterile; 00597001504 Femur CR PRC FEM COMP SIZE EMR Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.

Z-2009-2013

June 07, 2013

Class II

Product summary

Firm
Zimmer, Inc.
Event
Event 65413
Status
Terminated
Classification
Class II
Quantity
192,355
Official record key
device-enforcement:Z-2009-2013

Official wording

Reason: Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d

Code information: all codes

Distribution pattern: Worldwide Distribution-USA (nationwide) and the country of Japan.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d