Skip to content
Recall Observatory FDA recall evidence

Device product

Howmedica Osteonics Corp. Stryker Orthopaedics Specialty Triathlon Left and Right A//P Sizer Adjustment Housing PER File K2401. For use in knee prosthesis implantation. NON-STERILE

Z-0964-2013

January 24, 2013

Class II

Product summary

Firm
Stryker Howmedica Osteonics Corp.
Event
Event 64456
Status
Terminated
Classification
Class II
Quantity
4 units (total of 2 sets)
Official record key
device-enforcement:Z-0964-2013

Official wording

Reason: The standard Triathlon Femoral Stylus does not fit into the Specialty Sizers and the "R" and "L" markings on the Right Sizer are reversed.

Code information: Catalog Number: I-K2401KF00 (note each is uniquely identified/marked as IK2401KF0L and IK2401KF0R - Left and Right). Lot Codes: AP5W52 and AP7H19.

Distribution pattern: 1 Branch (NJ) with one consignee (DE).

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The standard Triathlon Femoral Stylus does not fit into the Specialty Sizers and the "R" and "L" markings on the Right Sizer are reversed.