Synchron LXi 725 Clinical System; P/N 107405S Product Usage: The Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.

Z-1034-2013

February 12, 2013

Class II

Product summary

Firm
Beckman Coulter Inc.
Event
Event 64332
Status
Terminated
Classification
Class II
Quantity
83 units
Official record key
device-enforcement:Z-1034-2013

Official wording

Reason: Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator belt pulley rubbing against the light shield can fall into the reaction vessels (RVs) and present a risk of erroneous test results.

Code information: Part # A15642; All Serial Numbers

Distribution pattern: Worldwide distribution: USA (nationwide) and countries of: Aurstria, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Djibouti, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Netherlands, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator belt pulley rubbing against the light shield can fall into the reaction vessels (RVs) and present a risk of erroneous test results.