Datascope System 98/98xt, CS 100/CS100i and CS 300 Intra-Aortic Balloon Pump. The Intra-Aortic Balloon Pump (IABP) is an electromechanical system used to inflate and deflate intra-aortic balloons in cardiovascular procedures.

Z-1523-2014

March 21, 2014

Class I

Product summary

Firm
Maquet Datascope Corp - Cardiac Assist Division
Event
Event 67815
Status
Terminated
Classification
Class I
Quantity
11,881 units (US 5183 and OUS 6698)
Official record key
device-enforcement:Z-1523-2014

Official wording

Reason: Potential mechanical failure of the fan assembly associated with the power supply. A fan assembly failure could result in the power supply overheating and cause the IABP to shut down without warning.

Code information: IABP Model Numbers System 98 0998-00-0446-xx 0998-UC-0446-xx System 98XT 0998-00-0479-xx 0998-UC-0479-xx CS100i 0998-UC-0446Hxx 0998-UC-0479Hxx CS100 0998-00-3013-xx 0998-UC-3013-xx CS300 0998-00-3023-xx 0998-UC-3023-xx

Distribution pattern: Worldwide Distribution - USA (nationwide) and approximately 104 countries.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    assembly failure