NeuViz 16 Multi-Slice CT Scanner System, Part number (PN): 989605858501, Model: NeuViz 16. This is a computed X-ray, Tomography system. The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles. The system is suitable for all patients

Z-1165-2013

March 04, 2013

Class II

Product summary

Firm
Philips And Neusoft Medical Systems Co., Ltd.
Event
Event 64783
Status
Terminated
Classification
Class II
Quantity
23 units
Official record key
device-enforcement:Z-1165-2013

Official wording

Reason: User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table was moved backward while the tabletop was controlled by another person from the gantry left control panel.

Code information: Serial numbers: N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120020, N16E120022, N16E120023, N16E120024, N16E120052, N16E120053, and N16E130002.

Distribution pattern: Nationwide distribution: USA including states of: California, Connecticut, North Carolina, Nebraska, Ohio, South Carolina, Texas, and Puerto Rico.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table was moved backward while the tabletop was controlled by another person from the gantry left control panel.