RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 799, 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care.

Z-1542-2014

April 18, 2014

Class II

Product summary

Firm
Med-Mizer, Inc.
Event
Event 67871
Status
Terminated
Classification
Class II
Quantity
1
Official record key
device-enforcement:Z-1542-2014

Official wording

Reason: All configurations of the SS Retractabed, Clinical Contour, made prior to June 7 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps them from separating from the frame. If the weld on the tube breaks, it can lead to bed collapse or the rollers coming out of the channel on th

Code information: Serial Number 2000

Distribution pattern: USA (nationwide)

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    All configurations of the SS Retractabed, Clinical Contour, made prior to June 7 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps them from separating from the frame. If the weld on the tube breaks, it can lead to bed collapse or the rollers coming out of the channel on th