Integra CUSA excel. CEM Nosecone for CUSA EXcel System. The CEM Nosecone is intended for use CUSA Excel/CUSA Excel + Ultrasonic Surgical Aspiration System, CUSA handpiece, and Covidien Force FX Electrosurgical Generator. Used in surgical procedures where combined ultrasonic dissection and electrosurgical coagulation is desired, either simultaneously or independently.

Z-0920-2013

January 07, 2013

Class II

Product summary

Firm
Integra LifeSciences Corp.
Event
Event 64318
Status
Terminated
Classification
Class II
Quantity
USA 6,190, EU 22,972 and ROW 288 units
Official record key
device-enforcement:Z-0920-2013

Official wording

Reason: There is a potential for erosion of the CUSA EXcel tip used with the Nosecone. Specifically, the potential for tip erosion occurs when using a CUSA CEM Nosecone if the side of the tip, rather than the annulus of the tip, is in contact with the tissue. This erosion can potentially lead to tip breakage. There were no reports of a patient injury associated with these complaints.

Code information: CEM Nosecone Catalogue No. C6623 Lot numbers: 1092855, 1093129, 1100436, 1100437, 1101217, 1101328, 1103393, 1104265, 1105067, 1110311, 1110892, 1112182, 1112183, 1114112, 1114655, 1115225, 1115499, 1115500, 1120771, 1121141, 1121142, 1121383, 1121627, 1122258, 1122259, 1123049 and 1130710. CEM Nosecone, Catalogue No. C6636: 1091431, 1091679, 1091680, 1093475, 1095103, 1095416, 1095417, 1100242, 1100434, 1100435, 1101094, 1101218, 1101219, 1103104, 1103395, 1103849, 1103849, 1103850, 1103852, 1103917, 1103918, 1103922, 1104257, 1104263, 1105071, 1110475, 1110476, 1110978, 1111490, 1111493, 1112187, 1112188, 1112212, 1113179, 1113180, 1113663, 1114357, 1114358, 1114659, 1114660, 1115039, 1115502, 1115504, 1115505, 1120776, 1120777, 1121167, 1121168, 1121169, 1121565, 1121566, 1121567, 1122344, 1123051, 1123052, 1123508, 1123511 and 1124053.

Distribution pattern: Worldwide Distribution - USA and the countries of Austria, Australia, Bangladesh, Belgium, Bulgaria, Brazil, Switzerland, Canada, China, Costa Rica, Czech Republic, Germany, Dominican Republic, Egypt, Emirates, Spain, Finland, France, Great Britian, Hong Kong, Hungary, Indonesia, Ireland, Israel, India, Italy, Japan, Korea (South), Kuwait, Lebanon, Sri Lanka, Latvia, Macedonia, Mexico, Malaysia, Netherlands, Philippines, Pakistan, Poland, Portugal, Qatar, Serbia, Saudi Arabia, Singapore, Slovenia, Thailand, Tunisia, Taiwan, United Arab, Viet Nam and South Africa.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a potential for erosion of the CUSA EXcel tip used with the Nosecone. Specifically, the potential for tip erosion occurs when using a CUSA CEM Nosecone if the side of the tip, rather than the annulus of the tip, is in contact with the tissue. This erosion can potentially lead to tip breakage. There were no reports of a patient injury associated with these complaints.