Skip to content
Recall Observatory FDA recall evidence

Device product

QuickDraw Venous Cannula with Insertion Kit. ThruPort Systmes. Models QD22 (22 Fr) and QD25 (25 Fr). The product is packaged sterile and non-pyrogenic in a sealed, peel-type pouch. Two pouches (cannula and accessories) go into one shelf carton. One shelf carton goes in one shipper carton. Use of the QuickDraw venous cannula is indicated for patients undergoing cardiopulmonary bypass. The QuickDraw venous cannula serves to drain non-oxygenated blood from the venae cavae or right atrium during cardiopulmonary bypass. The product is packaged sterile and nona¿ pyrogenic in a sealed, peel-type pouch.

Z-1542-2013

May 14, 2013

Class II

Product summary

Firm
Edwards Lifesciences, LLC
Event
Event 65173
Status
Terminated
Classification
Class II
Quantity
22,616
Official record key
device-enforcement:Z-1542-2013

Official wording

Reason: Edwards is recalling the QuickDraw Venous Cannula due to complaints of cannula separation during withdrawal of the device when the Percutaneous insertion method is used.

Code information: Model QD22: Lot No. 59049758, 59054488, 59067502, 59074170, 59081626, 59083736, 59091652, 59094887, 59114755, 59117451, 59134508, 59149651, 59154310, 59178579, 59201243, 59212118, 59222657, 59233636, 59233637, 59246684, 59246701, 59255899, 59255900, 59269439, 59284780, 59287725, 59296462, 59299605, 59315339, 59320575, 59334652, 59356129, 59360072, 59365198, 59365199, 59374009, 59384309, 59390552, 59394166, 59394167, 59421250, 59448841 Model QD25: Lot No. 59046082,59046847,59052393,59052395,59056270,59059029,59060798,59062681,59069772,59073141,59077234,59079573,59085168,59085169,59097289,59101637,59106005,59110720,59122491,59126281,59131923,59140133,59142266,59143453,59145027,59158524,59158579,59163164,59184252,59191107,59201246,59207044,59209886,59214360,59214361,59216991,59222658,59222659,59233651,59246685,59246703,59255901,59268625,59284781,59287728,59296463,59299610,59313057,59320582,59320583,59334656,59344191,59346468,59352608,59356135,59360084,59374014,59374015,59414791,59425355,59437699,59437700,59448853

Distribution pattern: Worldwide distribution: US Nationwide and country of Canada.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Edwards is recalling the QuickDraw Venous Cannula due to complaints of cannula separation during withdrawal of the device when the Percutaneous insertion method is used.