Device product

ACCU-CHEK Inform II Blood Glucose Monitoring System bar code scanner. Used with the ACCU-CHEK Inform II System to quantitatively measure glucose (sugar).

Z-1113-2013

March 06, 2013

Class II

Product summary

Firm
Roche Diagnostics Operations, Inc.
Event
Event 64460
Status
Terminated
Classification
Class II
Quantity
Both devices 4467
Official record key
device-enforcement:Z-1113-2013

Official wording

Reason: Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards.

Code information: part number 05060311001, 05060303001

Distribution pattern: Nationwide Distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0