Device product

ACIST Medical Systems Inc. The ACIST Angiographic Injection System is intended to be used for the controlled infusion of radiopaque contrast media for angiographic procedures.

Z-1145-2013

March 15, 2013

Class II

Product summary

Firm
Acist Medical Systems
Event
Event 64840
Status
Terminated
Classification
Class II
Quantity
6640 units
Official record key
device-enforcement:Z-1145-2013

Official wording

Reason: ACIST Medical Systems is issuing a voluntary recall notice for three specific lots of BT2000 Automated Manifold Kits. For these specific lots, ACIST has received field reports related to ingress of air into the manifold body. This has been detected by the Air Column Detect Sensor halting the injection, and normally replacement of the BT2000 Automated Manifold Kit is necessary in order to proceed w

Code information: 23884007, 23884008, 23884009

Distribution pattern: Nationwide Distribution including the states of AL, AK, AZ, AR, CA, CO, CT, DC, GA, HI, ID, IL, IN, IA, KS, KY, MD,MI, MN, MS, MO, NV, NH, NJ, NM, NY,NC, OR, OK, OH, PA, TN, TX, VA and WA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    ACIST Medical Systems is issuing a voluntary recall notice for three specific lots of BT2000 Automated Manifold Kits. For these specific lots, ACIST has received field reports related to ingress of air into the manifold body. This has been detected by the Air Column Detect Sensor halting the injection, and normally replacement of the BT2000 Automated Manifold Kit is necessary in order to proceed w