Skip to content
Recall Observatory FDA recall evidence

Device product

Extremity Medical - Disposable Orthopedic Surgical Instrument - 1.6 mm Guidewire - Reference Number 101-00006 - Non-Sterile. Contained in Extremity Medical Surgical Kits including CompressX, Hallu.X, IO Fix and XMCP. Used in the selection of screw size when used with appropriate depth gauge.

Z-0952-2013

January 10, 2013

Class II

Product summary

Firm
Extremity Medical LLC
Event
Event 64330
Status
Terminated
Classification
Class II
Quantity
238 units
Official record key
device-enforcement:Z-0952-2013

Official wording

Reason: The 1.6 mm Guidewire is undersized up to 2 mm in length. When used with the depth gauge, the measurement reading will not correspond with the desired screw length. The selected screw size based on the depth gauge reading will be about 2 mm longer than intended.

Code information: Catalog Number 101-00006; Lot Number 00921538.

Distribution pattern: Worldwide Distribution-USA including the states of Florida, Pennsylvania, New Hampshire, Oklahoma, California, Maryland, Arizona, Ohio, North Carolina, Texas, Virginia, Illinois, New York and Tennessee, and the countries of Australia, Belgium, Israel, Netherlands and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0