Acuson SC2000 volume imaging ultrasound systems at software version 3.0. Mfg name: Siemens Medical Solutions USA, Inc. Mountain View, CA 94043 Medical Device for human use. Ultrasound imaging system used to measure anatomical structures and calculation packages that provide information to the clinician for clinical diagnosis purposes.

Z-0052-2014

July 10, 2013

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc.
Event
Event 65812
Status
Terminated
Classification
Class II
Quantity
26 units
Official record key
device-enforcement:Z-0052-2014

Official wording

Reason: Software issue results in inaccurate 3D image may result in ablation in wrong location.

Code information: Model# 10433816 Software Version 3.0 Serial#'s: 401239 401189 401214 400416 400599 400600 401254 401260 401259 401161 400836 401111 401290 400641 401116 401236 401245 401241 401255 401361 400141 401108 401109 401213 400675 400882

Distribution pattern: Worldwide Distribution - US (nationwide) including the states of IL, PA, AZ, OR, OH, NY, CO, TX and CA., and the countries of France, Germany, Japan, Portugal, Saudi Arabia, Taiwan and Turkey.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    Software issue