Xoran MiniCAT CT Scanner

Z-0037-2014

July 27, 2010

Class II

Product summary

Firm
Xoran Technologies, Inc.
Event
Event 66337
Status
Terminated
Classification
Class II
Quantity
320
Official record key
device-enforcement:Z-0037-2014

Official wording

Reason: A description of the defect in the product or the manner in which the product fails to comply wvith an applicable Federal standard; The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evalu

Code information: Minicat

Distribution pattern: Worldwide Distribution - Nationwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A description of the defect in the product or the manner in which the product fails to comply wvith an applicable Federal standard; The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evalu