Skip to content
Recall Observatory FDA recall evidence

Device product

Xoran Minicat00202

Z-0038-2014

July 27, 2010

Class II

Product summary

Firm
Xoran Technologies, Inc.
Event
Event 66338
Status
Terminated
Classification
Class II
Quantity
320
Official record key
device-enforcement:Z-0038-2014

Official wording

Reason: The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evaluation of the hazards reasonably related to defect or the failure to comply with the Federal standard; An evaluation of the health ha

Code information: minicat

Distribution pattern: Worldwide Distribution - US nationwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evaluation of the hazards reasonably related to defect or the failure to comply with the Federal standard; An evaluation of the health ha