Device product
Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004
Z-0571-2014
Product summary
- Event
- Event 66754
- Status
- Terminated
- Classification
- Class II
- Quantity
- 24 systems
- Official record key
device-enforcement:Z-0571-2014
Official wording
Reason: When connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003, and 8.00.0004, and is configured with option "Strip Zeros from Result" set to "Yes" may result in sample error
Code information: Software alll versions prior to: v7.00.0006 and v8.00. Includes: v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004
Distribution pattern: Nationwide Canada and Britain
Derived failure modes
-
Unknown
When connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003, and 8.00.0004, and is configured with option "Strip Zeros from Result" set to "Yes" may result in sample error