Device product

20mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC2000 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial.

Z-0424-2014

April 04, 2013

Class II

Product summary

Firm
B. Braun Medical, Inc.
Event
Event 65356
Status
Terminated
Classification
Class II
Quantity
1,746,079 units
Official record key
device-enforcement:Z-0424-2014

Official wording

Reason: Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag. Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution. The misaligned cannula may result i

Code information: Material BC2000, Lot numbers 0061275133, 0061275135, 0061275136, 0061279143, 0061279144, 0061279145, 0061279146, 0061289059, 0061289060, 0061289061, 0061289062, 0061289063, 0061289064, 0061289065, 0061289066, 0061289067 and 0061299664.

Distribution pattern: Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag. Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution. The misaligned cannula may result i