COBAS INTEGRA 800 Analyzer with software 9864.C2 Cobas. COBAS Part Numbers 28122474001, 28122474692, 04559126001, 04589459970 corresponding to the Analyzer with or without closed tube sampling accessory A fully automated system for clinical chemistry analysis intended for the in vitro quantitative/qualitative determination of analytes in body fluids.

Z-0168-2014

September 16, 2013

Class II

Product summary

Firm
Roche Diagnostics Operations, Inc.
Event
Event 66239
Status
Terminated
Classification
Class II
Quantity
527 units
Official record key
device-enforcement:Z-0168-2014

Official wording

Reason: Roche COBAS INTEGRA 800 system, Software Version 9864.C2 does not perform necessary ISE Service Actions. Whenever the ISE rack is removed, the ISE counters in the software are reset to zero and the corresponding Service Actions are not conducted automatically by the system. Additionally, certain maintenance activities that have to be performed manually by the operator are not requested on the sof

Code information: software version 9864.C2

Distribution pattern: Nationwide Distribution and Puerto Rico

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Roche COBAS INTEGRA 800 system, Software Version 9864.C2 does not perform necessary ISE Service Actions. Whenever the ISE rack is removed, the ISE counters in the software are reset to zero and the corresponding Service Actions are not conducted automatically by the system. Additionally, certain maintenance activities that have to be performed manually by the operator are not requested on the sof