Skip to content
Recall Observatory FDA recall evidence

Device product

Abacus 3.1, Pharmacy Calculator Product Usage: Provide pharmacy calculations for automated compounding.

Z-1847-2014

May 28, 2014

Class II

Product summary

Firm
Baxter Corporation Englewood
Event
Event 68256
Status
Terminated
Classification
Class II
Quantity
42
Official record key
device-enforcement:Z-1847-2014

Official wording

Reason: Baxter Corporation is initiating a field correction following an upgrade to ABACUS Software v3.1, when Clinimix 4.25110 is used in an order, the Nutritional Summary and any labels containing energy information may display 1,000 times less protein than is actually present in the solution.

Code information: Product Code: 8300-0167, 8300-0168, 8300-0169

Distribution pattern: US Nationwide Distribution and the state of VA.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0