Device product
FB Reagent (REF 70562). FB reagent is an additional test used for revealing the results of some miniaturized biochemical test included in the API and Rapid ID 32 strips. The API and ID 32 strips product lines are a standardized system combining several biochemical tests, which enables group or species identifications of microorganisms.
Z-1476-2014
Product summary
- Firm
- BioMerieux SA
- Event
- Event 67820
- Status
- Terminated
- Classification
- Class II
- Quantity
- 33,853
- Official record key
device-enforcement:Z-1476-2014
Official wording
Reason: bioMerieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493) and FB reagent (REF 70562). The activity issue lead to a false negative result of some Rapid ID 32 strips and API biochemical tests using the ZYM B reagent or FB reagent
Code information: Codes - 1002717180, Exp. Date 5/30/2014, 1002737220, Exp. Date 6/13/2014, 1002778550, Exp. Date 8/9/2014, 1002782960, Exp. Date 8/15/2014, 1002792840, Exp. Date 8/15/2014, 1002834130, Exp. Date 8/22/2014, 1002877370, Exp. Date 8/22/2014
Distribution pattern: Worldwide Distribution-USA (nationwide) including DC and Puerto Rico, and the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WV, WA, WI, and WY and the country of Algeria, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Benin, Bolivia, Bosnia, Brazil, Brunel, Bulgaria, Burkina Faso, Cambodia, Cameroon, Canada, Central African Republic, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, Erythrea, France, Gabon, Georgia, Germany, Ghana, Greece, Guatemala, Haiti, Herzegovina, Honduras, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jamaica, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Mali, Mexico, Moldavia, Morocco, Mozambique, Myanmar, Netherlands, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Thailand, Trinidad & Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Derived failure modes
-
Unknown
bioMerieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493) and FB reagent (REF 70562). The activity issue lead to a false negative result of some Rapid ID 32 strips and API biochemical tests using the ZYM B reagent or FB reagent