Device product
Q-Station Quantification Software. Q-Station 3.0 with a2DQ and/or aCMQ Q-Apps are affected. Model: Q-Station, Catalog number: 795088, and Part Number: 989605382391. Intended to manage, view, analyze, and report qualitative and quantitative image data from ultrasound exams.
Z-0818-2015
Product summary
- Event
- Event 69790
- Status
- Terminated
- Classification
- Class II
- Quantity
- 43 units (13 in US and 30 outside US)
- Official record key
device-enforcement:Z-0818-2015
Official wording
Reason: When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ventricular Ejection Fraction (EF) may be higher, than the ESV and EF calculated by manual tracing without the use of QLAB. An incorrect EF calculation could lead to misdiagnosis and/or delayed or incorrect therapy.
Code information: SERIAL NUMBERS FOR FCO 79500321- Q-Station Quantification Software SYSTEM: 16B7-KADBT53D-15A9, 196B-UAD585A1-159E, 1ED3-QACA8669-1588, 2187-VAC5T6CD-157D, 2C46-QADA8FF0-15A8, 2EFA-RAD51054-159D, 3716-IAC49180-157C, 41D5-NADA1AA3-15A7, 4489-SAD49B07-159C, 49F1-IAC99BCF-1586, 4CA5-UAC41C33-157B, 5764-MAD9A556-15A6, 5A18-QAD425BA-159B, 5CCC-HACEA61E-1590, 5F80-MAC92682-1585, 6CF3-WAD93009-15A5, 725B-GACE30D1-158F, 750F-TAC8B135-1584, 8282-JAD8BABC-15A4, 8536-IAD33B20-1599, 87EA-PACDBB84-158E, 8A9E-RAC83BE8-1583, 9811-WAD8456F-15A3, 9D79-WACD4637-158D, AAEC-MADD4FBE-15AD, ADA0-PAD7D022-15A2, B054-UAD25086-1597, B308-JACCD0EA-158C, B5BC-RAC7514E-1581, C07B-KADCDA71-15AC, C897-TACC5B9D-158B, D60A-GADC6524-15AB, D8BE-KAD6E588-15A0, DB72-RAD165EC-1595, DE26-WACBE650-158A, E0DA-XAC666B4-157F, EE4D-IAD6703B-159F, F3B5-IACB7103-1589, F669-PAC5F167-157E, H944-VACAFBB6-1589, MBF8-WAC57C1A-157E, T3DC-SAD5FAEE-159F, and Y690-GAD07B52-1594.
Distribution pattern: Worldwide distribution (for FCO 83000190): US, including the states of AK, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NJ, NY, OH, OR, PA, SC, SD, TN, TX, VA, WA, and WI; and the countries of United Arab Emirates, Austria, Australia, Belgium, Canada, Switzerland, Colombia, Germany, Ecuador, Estonia, Spain, Finland, France, Great Britain, Hong Kong, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, Norway, New Zealand, Philippines, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Turkey, Taiwan, and Vietnam. ********************************************************************************** Worldwide Distribution -- US, including the states of KY, TN, ME, CA, PA, NC, CT, NY, SC, IL, IN, and WI; and the countries of Argentina, Spain, India, Italy, Austria, Germany, Belgium, Canada, Finland, Switzerland, Greece, Malaysia, Sweden, Australia, Netherlands, Mexico, Peru, Taiwan, Japan, France, and Norway.
Derived failure modes
-
Unknown
When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ventricular Ejection Fraction (EF) may be higher, than the ESV and EF calculated by manual tracing without the use of QLAB. An incorrect EF calculation could lead to misdiagnosis and/or delayed or incorrect therapy.