PKG, GRASPER FORCEPS, SERRSTED FENESTRATED, P/N 0250080318. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Z-0681-2015

November 17, 2014

Class II

Product summary

Firm
Stryker Endoscopy
Event
Event 69838
Status
Terminated
Classification
Class II
Quantity
180,573 units total
Official record key
device-enforcement:Z-0681-2015

Official wording

Reason: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Code information: 543367 544893 842088 845454 941849 942135 942844 943742 945442 1040419 1041098 1042499 1042929 1043777 1044526 1044527 1045033 1045372 1047071 1141219 1141973 1142134 1142913 1143973 1144935 1145348 1145796 1146957 1147046 1241906 1243270 1244187 1246179 1341468 1341557 1343701 1344283 1344526 1345529 1346236 1346680 1440403 1440910 1442176 1442809 1443395 1443652 646462

Distribution pattern: Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.