Device product
Mobiletta Mira Mobile X-ray System
Z-1875-2013
Product summary
- Event
- Event 65605
- Status
- Terminated
- Classification
- Class II
- Quantity
- 39
- Official record key
device-enforcement:Z-1875-2013
Official wording
Reason: Siemens issued a Field Safety Alert about the potential risk to users when operating the Mobilett Mira. When a Mobilett Mira is regularly exposed to extensive mechanical strain (driven over high thresholds for doors or elevators), it is possible that one of its main wheel shafts could break because of material fatigue, resulting in the chassis falling on the ground on the affected wheel side. This
Code information: Model number 10273100
Distribution pattern: US Distribution including the states of IA, SD, NC, WI, OH, NE, TX, KS, ID, MI, GA, AR and IL.
Derived failure modes
-
Unknown
Siemens issued a Field Safety Alert about the potential risk to users when operating the Mobilett Mira. When a Mobilett Mira is regularly exposed to extensive mechanical strain (driven over high thresholds for doors or elevators), it is possible that one of its main wheel shafts could break because of material fatigue, resulting in the chassis falling on the ground on the affected wheel side. This