Skip to content
Recall Observatory FDA recall evidence

Device product

Pediatric, 10Fr, Intermittent Urinary Catheters Model: P10 Pouch Label UPC (P10) *+M405P100C* Intermediate Box Label UPC (P10) *+M405P103F* Carton Label UPC (P10) *+M405P105H* The product is distributed individually or in intermediate boxes containing 30 individual, sterile packaged catheters (PA/PE film and dialyzing paper pouches). There are 30 catheters packaged in sterile pouches that are placed into an intermediate box, 10 of these intermediate boxes packaged into one case. Intended to be used by pediatric males and females for the purpose of bladder drainage

Z-0496-2015

November 11, 2014

Class II

Product summary

Firm
Cure Medical LLC
Event
Event 69737
Status
Terminated
Classification
Class II
Quantity
192 cases = 1,920 boxes = 57,600 each
Official record key
device-enforcement:Z-0496-2015

Official wording

Reason: Some of the intermediate boxes within lot 140417-3, were labeled as P8, a French Size 8 pediatric catheter but actually contain P10, French Size 10 pediatric catheters. Trying to use a P10 catheter for bladder drainage, where P8 catheters have been prescribed, may result in acute urinary retention; and, forced attempts to insert the French Size 10 catheter lead to urethral trauma.

Code information: Lot Number: 140417-3

Distribution pattern: Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Some of the intermediate boxes within lot 140417-3, were labeled as P8, a French Size 8 pediatric catheter but actually contain P10, French Size 10 pediatric catheters. Trying to use a P10 catheter for bladder drainage, where P8 catheters have been prescribed, may result in acute urinary retention; and, forced attempts to insert the French Size 10 catheter lead to urethral trauma.