ExacTrac is intended to be used to place patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, in order to treat lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. ExacTrac may also be used to monitor the patient position during the treatment.

Z-2146-2014

June 24, 2014

Class II

Product summary

Firm
Brainlab AG
Event
Event 68741
Status
Terminated
Classification
Class II
Quantity
15 systems
Official record key
device-enforcement:Z-2146-2014

Official wording

Reason: When using multiple isocenters (radiation treatment targets) within a single plan, in certain workflow conditions ExacTrac v.6.0.4 might move the patient to an unintended isocenter position, despite displaying the green "OK" icon. If this anomaly occurs and is not detected by the user, the radiation treatment dose at the linear accelerator may be delivered to the unintended target position.

Code information: ExacTrac Version 6.0.4; Model/Catalog Numbers: 1) 20833C EXACTRAC 6.0 IR POSITIONING SOFTWARE; 2) 20834C EXACTRAC 6.0 X-RAY POSITIONING SOFTWARE; 3) 20835C EXACTRAC 6.0 IR MONITORING SOFTWARE; 4) 20853C EXACTRAC 6.0 X-RAY LIMITED SOFTWARE; 5) 49926B ET UPGRADE SOFTWARE 3.X TO 6.0 (IR+XR); 6) 49927B ET UPGRADE SOFTWARE 4.X TO 6.0 (IR+XR); 7) 49928B ET UPGRADE SOFTWARE 5.X TO 6.0 (IR+XR); 8) 49933B ET UPGRADE TRUEBEAM 5.5 TO 6.0 LIMITED; 9) 49934A EXACTRAC UPGRADE 5.5 TO 6.0 BASIC; 10) 49936A ET SOFTWARE UPDATE 6.0.X TO 6.0.4; 11) 49938 ET SOFTWARE UPDATE 6.0.3TO 6.0.4; 12) 49939 ET SOFTWARE UPDATE 6.0.X to 6.0.4 P&R; 13) 49973B ET DATA PREP / REVIEW SYSTEM; 14) 49996B ET UPGRADE SOFTWARE X.X TO 6.0 SW ONLY; 15) 49997B ET UPGRADE SOFTWARE 5.5 TO 6.0 LIMITED; 16) 49998B ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT

Distribution pattern: Worldwide Distribution - USA including CA, CT, and IL; Internationally to France, Germany, Japan, Netherlands, Russia, and Slovenia.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    When using multiple isocenters (radiation treatment targets) within a single plan, in certain workflow conditions ExacTrac v.6.0.4 might move the patient to an unintended isocenter position, despite displaying the green "OK" icon. If this anomaly occurs and is not detected by the user, the radiation treatment dose at the linear accelerator may be delivered to the unintended target position.