Device product

cobas b 123 Fluid Pack COOX REF 05169992001 200 Product Usage: The cobas b 123 POC system is a fully automated POC system for whole blood in vitro measurement of pH, blood gases (B3G), electrolytes NAT, K+, iCaWt (ISE), hematocrit (THct), metabolites (GLu, Lac) total hemoglobin (tHb), hemoglobin derivatives (O2Hb, HHb, COHb, MetHb), and oxygen saturation (SO2). The cobas b 123 POC system calculates derived parameters. Depending on the equipment configuration of the instrument, the Sensor Cartridge and the Fluid Pack used, the following parameters are measured in human whole blood and QC materials.

Z-2446-2015

July 14, 2015

Class II

Product summary

Firm
Roche Diagnostics Operations, Inc.
Event
Event 71786
Status
Terminated
Classification
Class II
Quantity
82 fluid packs
Official record key
device-enforcement:Z-2446-2015

Official wording

Reason: Low PO2 results. QC failures of the PO2 parameter, affecting primarily Levels 1 and 2, caused by a calibration issue with the PO2 parameter. This issue may not be detected since QC results can be below mean values, but still within 2 standard deviations (SD) limits. Potential for erroneously low PO2 results in patient samples, especially in blood samples with PO2 values below 50 mmHg

Code information: cobas b 123 Fluid Pack COOx Catalog 05169992001, lot numbers 21446113 (exp 8/8/2015), 21446123 (exp 8/18/2015), 21446153 ( exp 9/18/2015), 21446183 (exp 9/19/2015), and 21446203 (exp 10/01/2015)

Distribution pattern: US Nationwide Distribution in the states of LA, SC, CA, TX, NE, VA and WV including: Puerto Rico.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Low PO2 results. QC failures of the PO2 parameter, affecting primarily Levels 1 and 2, caused by a calibration issue with the PO2 parameter. This issue may not be detected since QC results can be below mean values, but still within 2 standard deviations (SD) limits. Potential for erroneously low PO2 results in patient samples, especially in blood samples with PO2 values below 50 mmHg