Device product
VERO Linear Accelerator System, Model Number MHI-TM2000; Common Name: Medical Linear Accelerator. MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.
Z-2653-2014
Product summary
- Event
- Event 68909
- Status
- Terminated
- Classification
- Class II
- Quantity
- 3 units in use in the US
- Official record key
device-enforcement:Z-2653-2014
Official wording
Reason: Software Anomaly: If a user changes the calendar setting from Workday to Holiday or vice versa, that would erroneously change the status of treatment/fractions, which are completed and to become Treated (completed) status on the exact day when such change is made, to Untreated status. Furthermore, such treatment would be cloned and mistakenly added to the schedule as Untreated treatment
Code information: Software Version 3.0.0 and after. Serial Numbers 201902, 203901, and 203919
Distribution pattern: Distributed in the states of NY, FL, and TX.
Derived failure modes
-
Device software or design
Software Anomaly