Device product

The Brilliance 64 and Ingenuity CT are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.

Z-2664-2014

September 01, 2011

Class II

Product summary

Firm
Philips Medical Systems (Cleveland) Inc
Event
Event 69045
Status
Terminated
Classification
Class II
Quantity
6
Official record key
device-enforcement:Z-2664-2014

Official wording

Reason: During a Field Test (a customer external evaluation period), the customer detected a problem in the CIRS 4.0 Beta 2 software. When using the CIRS 4.0 Beta 2 software, during the recon during ready stage of reconstruction, images may be overlapped with, or superimposed on other images. Philips has disabled this feature at all sites evaluating 4.0 Beta 2 software. Philips will issue a software up

Code information: 91003, 30018, 95504, 300004, 300030, 300010, 95130, 95157

Distribution pattern: Worldwide Distribution: US distribution in OH and countries of: Australia, China, Belgium, Israel, and Sweden.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software. Philips will issue