Skip to content
Recall Observatory FDA recall evidence

Device product

iba COMPASS SW Version 3.0, Catalog Number CS10-100 (medical linear accelerator) radiological

Z-0456-2014

August 20, 2013

Class II

Product summary

Firm
Iba Dosimetry Gmbh
Event
Event 66808
Status
Terminated
Classification
Class II
Quantity
33 units
Official record key
device-enforcement:Z-0456-2014

Official wording

Reason: The development partner of IBA informed that there is problem in a SW component used in COMPASS when converting Region of Interest (ROI) geometries to dose grid representation. Under certain conditions this problem generates too large ROI voxel volumes causing erroneous dose statistics and DVHs.

Code information: SW Version 3.0. Catalong Number CS10-100

Distribution pattern: Worldwide Distribution - US Distribution including the states of CT, MA, OH, CA, SC, NY and OK, and the countries of International distribution (from Germany) Poland, Germany, GB, The Netherlands, Spain, Turkey, Canada, Mexico, China, Hong Kong and Singapore.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The development partner of IBA informed that there is problem in a SW component used in COMPASS when converting Region of Interest (ROI) geometries to dose grid representation. Under certain conditions this problem generates too large ROI voxel volumes causing erroneous dose statistics and DVHs.