Sensei X Robotic Catheter System, model numbers 02057 and 08257. - Set up joint (SUJ)mechanical clamp that secures the SUJ to the catheter laboratory table. Product Usage: The Sensei X Robotic Catheter System is an electromechanical software controlled system designed to facilitate navigation and control of Hansen Medicals robotically steerable catheters for collecting electrophysiological data within the heart atria with electro-anatomic mapping and recording systems.

Z-0896-2015

December 04, 2014

Class II

Product summary

Firm
Hansen Medical Inc
Event
Event 69855
Status
Terminated
Classification
Class II
Quantity
59 USA, 40 OUS
Official record key
device-enforcement:Z-0896-2015

Official wording

Reason: The Set Up Joint subassembly of the Sensei X Robotic Catheter System may become loose relative to the operating table with the potential for the device to fall causing, uncontrolled catheter movement. cardiac tampenade, delayed procedure, or crush injury to the patient and/or user.

Code information: Model numbers 02057 and 08257, all installed Sensei X Robotic systems are subject to this action.

Distribution pattern: Worldwide Distribution - US Nationwide and the countries of Italy, Czech Republic, Russia, Spain, Singapore, France, Germany, Denmark, Netherlands and Australia.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The Set Up Joint subassembly of the Sensei X Robotic Catheter System may become loose relative to the operating table with the potential for the device to fall causing, uncontrolled catheter movement. cardiac tampenade, delayed procedure, or crush injury to the patient and/or user.