Device product

Siemens Healthcare Diagnostics Dimension Vista Systems, used with the Dimension Vista Intelligent Lab System software version 3.6.1. In vitro diagnostic analyzer.

Z-2054-2015

May 21, 2015

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 71315
Status
Terminated
Classification
Class II
Quantity
2315
Official record key
device-enforcement:Z-2054-2015

Official wording

Reason: Two software issues may occur in Vista software version 3.6.1. Issue #1: Samples may stop processing without notification. Issue #2: Timing issue in Vista 1500 may cause wrong reagent or no reagent delivery. This may lead to unflagged, unexpected low results and results flagged with assay errors.

Code information: Siemens Material Number (SMN): 10284473, 10488224, 10444801, 10444802

Distribution pattern: Worldwide Distribution - Nationwide Distribution including Puerto Rico., and the countries of : Austria, Australia, Belgium, Canada, Denmark, France, Finland, Germany, Japan, Italy, New Zealand, Netherlands, Czech Republic, Portugal, Republic Korea, Norway, Spain, Switzerland, Slovakia, Saudi Arabia, Slovenia, and United Kingdom,

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software issues may occur in Vista software version 3.6.1. Issue