Device product
Siemens Healthcare Diagnostics Dimension Vista Systems, used with the Dimension Vista Intelligent Lab System software version 3.6.1. In vitro diagnostic analyzer.
Z-2054-2015
Product summary
- Event
- Event 71315
- Status
- Terminated
- Classification
- Class II
- Quantity
- 2315
- Official record key
device-enforcement:Z-2054-2015
Official wording
Reason: Two software issues may occur in Vista software version 3.6.1. Issue #1: Samples may stop processing without notification. Issue #2: Timing issue in Vista 1500 may cause wrong reagent or no reagent delivery. This may lead to unflagged, unexpected low results and results flagged with assay errors.
Code information: Siemens Material Number (SMN): 10284473, 10488224, 10444801, 10444802
Distribution pattern: Worldwide Distribution - Nationwide Distribution including Puerto Rico., and the countries of : Austria, Australia, Belgium, Canada, Denmark, France, Finland, Germany, Japan, Italy, New Zealand, Netherlands, Czech Republic, Portugal, Republic Korea, Norway, Spain, Switzerland, Slovakia, Saudi Arabia, Slovenia, and United Kingdom,
Derived failure modes
-
Device software or design
software issues may occur in Vista software version 3.6.1. Issue