Skip to content
Recall Observatory FDA recall evidence

Device product

VITROS 4600 Chemistry System, Catalog Number 6802445, Unique Device Identifier Number 10758750012343, Software Version 3.2 and below; IVD. Product Usage: The VITROS 4600 Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.)

Z-0642-2016

November 03, 2015

Class II

Product summary

Firm
Ortho-Clinical Diagnostics
Event
Event 72597
Status
Terminated
Classification
Class II
Quantity
Domestic: 157, Foreign: 270
Official record key
device-enforcement:Z-0642-2016

Official wording

Reason: Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems continue to operate and process results without notification to the User, possibly producing erroneous results.

Code information: Software Version 3.2 and below; Serial Numbers J46001001 -J4600636

Distribution pattern: Worldwide Distribution - US Nationwide and the countries of Puerto Rico, and foreign distribution to: Argentina, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Israel, Italy, Japan, Kenya, Mexico, Panama, Philippines, Poland, Romania, Russia, Singapore, Spain, Taiwan, Turkey, and Venezuela.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    Software anomaly