Skip to content
Recall Observatory FDA recall evidence

Device product

Central Line >34 Weeks Kit. Cardiology department. The Central Line >34 Weeks Kit (DYNDC1987A) is intended for use on patients greater than 34 weeks of age. The central venous catheter, also called a central line, is a long, thin, flexible tube used to give medicines, fluids, nutrients, or blood products over a long period of time, usually several weeks or more. A catheter is often inserted in the arm or chest through the skin into a large vein. The catheter is threaded through this vein until it reaches a large vein near the heart.

Z-2458-2015

August 03, 2015

Class II

Product summary

Firm
Medline Industries Inc
Event
Event 71889
Status
Terminated
Classification
Class II
Quantity
320 units for lot 14DB9277 & 320 units for lot 14PB2878
Official record key
device-enforcement:Z-2458-2015

Official wording

Reason: The print plate artwork for this lot was found to state: Central Line <34 Weeks Kit displaying an incorrect sign in front of the 34. The description for this device should have stated Central Line >34 Weeks Kit. The defect may lead to the product to be used on patient population that is <34 weeks rather than intended population of >34 weeks.

Code information: Central Venous Catheter Kit for patients >34 weeks old with the coding below were affected. Lot Numbers: 14DB9277 and 14PB2878 Unit No: DYNDC1987A Expiration Date: 06/2015 Serial No: N/A UPC code: N/A

Distribution pattern: US including PA

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The print plate artwork for this lot was found to state: Central Line <34 Weeks Kit displaying an incorrect sign in front of the 34. The description for this device should have stated Central Line >34 Weeks Kit. The defect may lead to the product to be used on patient population that is <34 weeks rather than intended population of >34 weeks.