Various Lots of Expel Drainage Catheter with TwistLoc Hub Biliary Drainage Catheter and Biliary Drainage Catheter Kits

Z-1355-2015

February 25, 2015

Class II

Product summary

Firm
Boston Scientific Corporation
Event
Event 70604
Status
Terminated
Classification
Class II
Quantity
289
Official record key
device-enforcement:Z-1355-2015

Official wording

Reason: Complaints of device fragmentation after the catheter was implanted in the biliary system. The most serious patient risk for this issue is additional intervention for fragment retrieval using minimally invasive methods and, in some of the reported cases, this was performed to retrieve the device fragments.

Code information: UPN # LOT # H7493933108200 17245003, 17416605, 17489192 H7493933108350 17245004, 17245005, 17370709, 17408756 H749393310835K1 17411333, 17493554 H7493933110200 17318047 H7493933110350 17275822, 17438003, 17494025, 17501572 H749393311035K1 17328808, 17403871, 17416578, 17459144 H7493933110400 17275824, 17416608 H7493933112350 17245006, 17259580, 17275819 H7493933114350 17275815, 17338207 with Expiration Dates between January 30, 2016 and May 8, 2016

Distribution pattern: Worldwide Distribution - US Distribution to the states of : CA, WI, FL and OH., and to the countries of: Belgium, Czech Republic, Finland, France, Germany, Great Britain, Italy, Spain, Switzerland, Austria, Hungary, Netherlands and Portugal.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Complaints of device fragmentation after the catheter was implanted in the biliary system. The most serious patient risk for this issue is additional intervention for fragment retrieval using minimally invasive methods and, in some of the reported cases, this was performed to retrieve the device fragments.