Skip to content
Recall Observatory FDA recall evidence

Device product

Model BP7507 OPTI LION E-Plus Cassettes 25 Cassettes with Sample Probes, OPTIMedical Cassettes are consumable used in the OPTI LION Electrolyte Analyzer intended to be used for the measurement of sodium, potassium, chloride, ionized calcium and pH in samples of whole blood, serum, and plasma.

Z-0029-2014

August 09, 2013

Class II

Product summary

Firm
OPTI Medical Systems, Inc
Event
Event 66105
Status
Terminated
Classification
Class II
Quantity
361 boxes of 25 cassettes each
Official record key
device-enforcement:Z-0029-2014

Official wording

Reason: OPTI LION E-Plus Cassettes were labeled with an incorrect barcode. The incorrect barcode could perfect the performance of one or more parameters as the product ages. Also a Customer complaint from a distributor reported that a customer was getting QC failures.

Code information: Lots 252810, 315815, 319810, 319815, 321817, 367812, 370812

Distribution pattern: Worldwide distribution: US (nationwide) including states of: CA, MA, MN, and VA; and countries of: China, Guatemala, India, Iraq, South Korea, and Untied Arab Emirates.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    OPTI LION E-Plus Cassettes were labeled with an incorrect barcode. The incorrect barcode could perfect the performance of one or more parameters as the product ages. Also a Customer complaint from a distributor reported that a customer was getting QC failures.