Skip to content
Recall Observatory FDA recall evidence

Device product

BZO Strip Dip Card (300ng/ml)*, Item No. 100170

Z-0490-2016

November 16, 2015

Class III

Product summary

Firm
Ameditech Inc
Event
Event 72651
Status
Terminated
Classification
Class III
Quantity
31,472 kits
Official record key
device-enforcement:Z-0490-2016

Official wording

Reason: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

Code information: 141233 141669 141790 141820 141958 142082 142531 142532 142835 143020 143471 143553 152021 152432 152546 153094 153538 152104

Distribution pattern: Nationwide Distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.