EndoVive 3s Low Profile Balloon Kits Part Number: M00548870 (XMD P/N 70-0050-620) The EndoVive 3s Low Profile Balloon is intended to provide gastric access through a gastrointestinal stoma tract for enteral feeding, medication administration and gastric decompression. The EndoVive 3s Low Profile Balloon is indicated for use in adult and pediatric patients who require long-term feeding, are unable to tolerate oral feeding, are at low risk for aspiration or require gastric decompression and/or medication delivery directly into the stomach.

Z-0951-2016

December 23, 2015

Class II

Product summary

Firm
Xeridiem Mediem Medical Devices Inc
Event
Event 73024
Status
Terminated
Classification
Class II
Quantity
37
Official record key
device-enforcement:Z-0951-2016

Official wording

Reason: A feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric contents to leak out of the device in some instances.

Code information: Part Number: M00548870 (XMD P/N 70-0050-620) Lot Number: 929729

Distribution pattern: Nationwide distribution to MA only.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric contents to leak out of the device in some instances.