Device product
Assay Data Disk (ADD), All DRV versions that support product within expiry dating Automates pre-analytical and post-analytical sample and data management for in vitro diagnostic use. VITROS Urine Assays: VITROS Calcium (Ca), Magnesium (Mg), and Phosphorus (Phos) requires acidification pretreatment prior to urine assay testing
Z-1729-2015
Product summary
- Event
- Event 70918
- Status
- Terminated
- Classification
- Class II
- Quantity
- 2,387 units Total (987 domestically & 1400 internationally)
- Official record key
device-enforcement:Z-1729-2015
Official wording
Reason: Software anomaly regarding urine samples that require acidified pretreatment. The software has allowed urine samples for assays that require acidified pretreatment to be metered from the same sample container as assays that required no pretreatment when the sample was programmed.
Code information: Product Code: 6801876 Unique Device Identifier Numbers: 10758750001576
Distribution pattern: Worldwide Distribution - US Nationwide, Brazil, Chile, Mexico, Australia, Hong Kong, India, Taiwan, Belgium, Denmark, France, Ireland, Italy, Portugal, Romania, Spain, Sweden and the UK.
Derived failure modes
-
Device software or design
Software anomaly