Device product
SIEMENS brand ARTISTE Linear Accelerator Linac systems with component: any photon unflat beam option (Multiple X feature) in combination with the IMRT or mARC option. A family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. The ARTISTE/ONCOR systems are manufactured by: SIEMENS AG, MEDICAL SOLUTIONS, GERMANY and distributed SIEMENS Healthcare IM CR RO, CR Radiation Oncology, Martinez, CA
Z-0438-2014
Product summary
- Event
- Event 66724
- Status
- Terminated
- Classification
- Class II
- Quantity
- 26 units
- Official record key
device-enforcement:Z-0438-2014
Official wording
Reason: A safety risk regarding a potentially existing dark current radiation phenomenon on the Linear Accelerator (LINAC) in combination with IMRT or mARC (rotational IMRT) treatments using unflat (flattening filter free) beams. In case there are many long periods of field shape changes in a treatment plan, the dose due to dark current radiation may sum up to a significant value.
Code information: Serial Numbers: 5364, 5419, 5640, 5737, 5823, 5830, 5166, 5342, 5416, 5444, 5479, 5485, 5628, 5745, 5746, 5762, 5763, 5764, 5772, 5781, 5783, 5797, 5815, 5819, 5820, 5822, 5829, 5837, 5839, 5888, 5896, 5911, 5937
Distribution pattern: Worldwide Distribution: USA (nationwide) including states of: FL, NC, NY, PA, TN, UT, and WI; and countries of: Angola, Brazil, Canada, P.R. China, Germany, India, Japan, Republic Korea, Poland, and Thailand.
Derived failure modes
-
Unknown
A safety risk regarding a potentially existing dark current radiation phenomenon on the Linear Accelerator (LINAC) in combination with IMRT or mARC (rotational IMRT) treatments using unflat (flattening filter free) beams. In case there are many long periods of field shape changes in a treatment plan, the dose due to dark current radiation may sum up to a significant value.