Skip to content
Recall Observatory FDA recall evidence

Device product

Terumo Custom Cardioplegia Delivery Set - Heat Exchanger Cardiopulmonary Bypass Item Numbers: 5852, 16010 Product Usage: This is a Cardioplegia Delivery System that is designed to delivery blood/solution to the patient undergoing cardiopulmonary bypass surgery. The set is designed to deliver blood/solutions at a 4:1 ratio. In includes a Conducer Heat Exchanger and a Bubble Trap module. Additionally, the set is constructed with a female luer connector that is the subject of the correction/removal activity

Z-0156-2015

March 01, 2011

Class II

Product summary

Firm
Terumo Cardiovascular Systems Corp
Event
Event 58185
Status
Terminated
Classification
Class II
Quantity
596 units; 876 units
Official record key
device-enforcement:Z-0156-2015

Official wording

Reason: Defect in the luer connector can compromise the seal on the female side of the connector and the compromised seal can introduce air in the table line

Code information: Item Number: 58520-Lot Numbers: MF24, MM11, MN01. Manufacturing dates May 24, 20 I 0 through November I, 2010 Item Number: 1610-Lot Numbers: MM11, MN08, MNI5, MN29 MP07. Manufacturing dates October 11, 20 I 0 through December 7, 20 I 0

Distribution pattern: Worldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Defect in the luer connector can compromise the seal on the female side of the connector and the compromised seal can introduce air in the table line