Device product
Inspiration Ventilator System, i Series and LS 12.1 LCD. The Inspiration 7i / 5i Ventilator System is to be used by healthcare professionals in hospitals or healthcare facilities and intra-hospital transport.
Z-0230-2016
Product summary
- Event
- Event 72491
- Status
- Terminated
- Classification
- Class I
- Quantity
- 6441 units total (251 units in US)
- Official record key
device-enforcement:Z-0230-2016
Official wording
Reason: A component on the systems power board can fail in a manner which can cause the ventilator to shut down without alarming.
Code information: All models manufactured prior to 2015 are affected
Distribution pattern: Worldwide Distribution - US (Nationwide) and Internationally to: Algeria Argentina Australia Austria Bahrain Bangladesh Barbados Belarus Belgium Bolivia Canada Chile China Colombia Czech Rep Ecuador Egypt England Estonia Finland Germany Ghana Gibraltar Greece Guatemala Holland Hong Kong Hungary India Indonesia Iran Iraq Ireland Israel Italy Japan Jordan Kenya Korea Kuwait Laos Latin America Lebanon Libya Lithuania Malaysia Mexico Morocco Nepal Netherlands New Zealand Norway Pakistan Palestine Panama Peru Philippines Poland Portugal Russia Saudi Arabia Serbia Singapore South Africa Spain Sri Lanka Sudan Switzerland Syria Taiwan Thailand Turkey UAE Ukraine USA Venezuela Vietnam Yemen Zimbabwe
Derived failure modes
-
Unknown
A component on the systems power board can fail in a manner which can cause the ventilator to shut down without alarming.