Device product
BG cartridges for IRMA TRUPOINT Blood Gas Analyzer, PN 048103. BG Cartridge are single use, disposable, in vitro diagnostics intended for point of care (POC), professional, use with the IRMA TRUpoint¿ Blood Analysis System.
Z-0160-2016
Product summary
- Firm
- LifeHealth, LLC
- Event
- Event 72305
- Status
- Terminated
- Classification
- Class II
- Quantity
- 1735 boxes of 25. (43375 Cartridges)
- Official record key
device-enforcement:Z-0160-2016
Official wording
Reason: May report values outside the published performance specifications for pH. After a certain point in shelf life, the pH sensor may give results that are falsely elevated. The discrepancy is greater as shelf life progresses and can be seen on the aqueous controls before the blood.
Code information: EOSCU EOTUZ EOVMW EOYUL EPANW EPCED EPDVM EPFOJ EPIXX EPKOJ EPNXA
Distribution pattern: US: Nationwide (AL, AR, CA, CO, FL, IL, IN, IA, KS, KY, MD, MA,MI, MN, MS, MO, NE, NJ, NY,NC, OH, OK, PA, TN, TX, VA) OUS: ALBANIA, BELGIUM, BOLIVIA, CANADA, CHILE, CHINA, DENMARK. DOMINICAN REPUBLIC, ECUADOR, EGYPT, ENGLAND, ESTONIA, PHILIPPINES, FRANCE, GERMANY, GREECE, ICELAND, INDONESIA, IRELAND, ITALY, JAMAICA, JAPAN, MOROCCO, NETHERLANDS, NORWAY, PERU, PORTUGAL, SOUTH AFRICA, SOUTH KOREA, SPAIN, TAIWAN, TUNISIA, TURKEY, UNITED ARAB EMIRATES, VIETNAM.
Derived failure modes
-
Unknown
May report values outside the published performance specifications for pH. After a certain point in shelf life, the pH sensor may give results that are falsely elevated. The discrepancy is greater as shelf life progresses and can be seen on the aqueous controls before the blood.