Device product

ProScreen 10 Panel Dip Card, Item No. PSD-10MOX, PSD-10MOB, PSD-10MMO300

Z-0518-2016

November 16, 2015

Class III

Product summary

Firm
Ameditech Inc
Event
Event 72651
Status
Terminated
Classification
Class III
Quantity
2,452 kits
Official record key
device-enforcement:Z-0518-2016

Official wording

Reason: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

Code information: 141963 142196 143166 143475 143827 152038 152172 152218 152323 152393 152394 152870 153362

Distribution pattern: Nationwide Distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.