Device product
ImmuTest 12 Panel Dip Card, Item No. IMD-12MO3
Z-0511-2016
Product summary
- Firm
- Ameditech Inc
- Event
- Event 72651
- Status
- Terminated
- Classification
- Class III
- Quantity
- 2,693 kits
- Official record key
device-enforcement:Z-0511-2016
Official wording
Reason: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
Code information: 144009 144078 144099 144125 144193 144268 152996 153077
Distribution pattern: Nationwide Distribution.
Derived failure modes
-
Unknown
Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.