Volcano s5i/s5ix/s5iz/CORE Intravascular Ultrasound Imaging and Pressure System; Product/Part numbers: s5i -806300-020 ; s5ix/s5iz - 807400-001; CORE - 400-0 I 00.02. Intravascular Imaging and Pressure System for evaluation of blood flow, perfusion and vascular structures.

Z-2044-2015

September 23, 2013

Class II

Product summary

Firm
Volcano Corp
Event
Event 71404
Status
Terminated
Classification
Class II
Quantity
227 total units - all models
Official record key
device-enforcement:Z-2044-2015

Official wording

Reason: A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR) within the Volcano System when the system is connected to the Xper Flex Cardia system via its aortic pressure auxiliary output.

Code information: Product/Part numbers: s5i -806300-020 ; s5ix/s5iz - 807400-001; CORE - 400-0 I 00.02. All software versions

Distribution pattern: Worldwide Distribution - US Nationwide and the countries of Belgium, United Kingdom, Denmark, Netherlands, Sweden, Iceland, France, Saudi Arabia, Australia, Philippines and Thailand.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR) within the Volcano System when the system is connected to the Xper Flex Cardia system via its aortic pressure auxiliary output.