Device product
PKG, 5MM X 45CM, INSERT, KELLY FORCEPS, P/N 0250080753 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Z-0746-2015
Product summary
- Event
- Event 69838
- Status
- Terminated
- Classification
- Class II
- Quantity
- 180,573 units total
- Official record key
device-enforcement:Z-0746-2015
Official wording
Reason: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Code information: 844129 940132 943736 944432 1041619 1041989 1044112 1044113 1044417 1045046 1045047 1141783 1142150 1142921 1143798 1241902
Distribution pattern: Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Derived failure modes
-
Unknown
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods