Recall Observatory FDA recall evidence

Device product

2.3MM Tapered Router, Product Number 5407-FA2-023 A router is a cutting accessory used in the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/Otology /Neurotology/ Otorhinolaryngology; Craniofacial (bones of the skull and supraorbital region); and Sternotomy. They are intended to be used with the Stryker CORE" system.

Z-2238-2016

June 24, 2016

Class II

Product summary

Firm
Stryker Instruments Div. of Stryker Corporation
Event
Event 74639
Status
Terminated
Classification
Class II
Quantity
3,163
Official record key
device-enforcement:Z-2238-2016

Official wording

Reason: A variation in flute depth on the routers was observed. The variation in flute depth may cause the routers to be more susceptible to breakages. Risk to Health: Injury to critical soft tissue requiring surgical intervention may occur.

Code information: Lot numbers: 15077027, 15077067, 15077097, 15078087, 15105017, 15108017, 15108047, 15108077, 15108087, 15108097, 15146017, 15146027, 15146037, 15146047

Distribution pattern: Worldwide Distribution - US (Nationwide) Netherlands, Canada, Japan, Switzerland, and United Kingdom

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A variation in flute depth on the routers was observed. The variation in flute depth may cause the routers to be more susceptible to breakages. Risk to Health: Injury to critical soft tissue requiring surgical intervention may occur.