Skip to content
Recall Observatory FDA recall evidence

Device product

ASP-1020 LipoTower with Aspiration Pump. Medical use.

Z-2821-2015

June 12, 2015

Class II

Product summary

Firm
MicroAire Surgical Instruments, LLC
Event
Event 72106
Status
Terminated
Classification
Class II
Quantity
3 LipoTower with Aspiration Pump
Official record key
device-enforcement:Z-2821-2015

Official wording

Reason: There is potential risk of electrical shock if lead wires inside the device become detached and the ground connection fails.

Code information: Serial numbers 14031102, 15021105, and 14101103.

Distribution pattern: Nationwide Distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is potential risk of electrical shock if lead wires inside the device become detached and the ground connection fails.