Reagent Management System (RMS) Dimension RxL Max HM STM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension¿ clinical chemistry system. The RMS doubles the available storage of refrigerated Flex¿ reagent cartridges on the Dimension¿ system from 44 to 88. The RMS loads, hydrates, and automatically removes Flex¿ reagent cartridges, allowing the Dimension¿ system to process samples without interruption.

Z-0461-2016

November 10, 2015

Class II

Product summary

Firm
Siemens Healthcare Diagnostics Inc.
Event
Event 72738
Status
Terminated
Classification
Class II
Quantity
151
Official record key
device-enforcement:Z-0461-2016

Official wording

Reason: Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind the RMS Waste Container.

Code information: Dimension RXL MAX HM STM Catalog Numbers 752000.941 (Domestic)/752000.951 (Intl). RMS for RxL Catalog Numbers 753000.901 (Domestic)/753000.902 (Intl). All serial numbers.

Distribution pattern: US Nationwide Distribution: FL, NY, SC, TX, LA, AZ, WI, IA, IL, OH, AL, AK, NH, OK, MD, CA, NM, MS, KY, NJ, MN, GA, CO, MA, SD, TN, MO, PA, MT, SC, NE, KS, VA, AL, NV, WY, OR, PR, MO, OK, GU, HI, AR, ME, VT, CT, NC and the countries of Canada and Mexico.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind the RMS Waste Container.