Skip to content
Recall Observatory FDA recall evidence

Device product

Total Temporomandibular Joint Replacement System- 50mm Rt Narrow Ti Mand Reconstruction of the temporomandibular joint

Z-0515-2015

October 17, 2014

Class II

Product summary

Firm
Biomet Microfixation, LLC
Event
Event 69644
Status
Terminated
Classification
Class II
Quantity
42
Official record key
device-enforcement:Z-0515-2015

Official wording

Reason: Laser etching on the parts is wider and deeper than the conditions previously validated.

Code information: Model 01-6550TI, Lot #s: 531890A, 531890B, 531900A, 531900B, 544220B, 544230A, 544230B, 544200A, 544200B, 544210A, 544210B, 544220A.

Distribution pattern: Worldwide Distribution - USA including PA, MN, CA and Internationally to Australia, Netherlands, France, and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Laser etching on the parts is wider and deeper than the conditions previously validated.