Device product
Black Mamba Suture Passer
Z-0471-2016
Product summary
- Firm
- Biomet, Inc.
- Event
- Event 72829
- Status
- Terminated
- Classification
- Class II
- Quantity
- 50
- Official record key
device-enforcement:Z-0471-2016
Official wording
Reason: There may be a separation and an insufficient weld between the spring and washer. The weld between the spring and washer may fracture, causing the Mamba Suture Passer instruments jaw to not actuate properly. 6 complaints have been received in which 3 have been reported with a delay up to 10 minutes.
Code information: Catalog Number: 110010849 Lot Number Identification: 167880, 169620, 253190, 520810, 538490, 551550, 716270, 960180
Distribution pattern: FL, NJ, WI, UT, TX, KY, KS, OH, GA, LA, AL, NY, NC, AR, MO, VA, CA, MN, IN, HI
Derived failure modes
-
Unknown
There may be a separation and an insufficient weld between the spring and washer. The weld between the spring and washer may fracture, causing the Mamba Suture Passer instruments jaw to not actuate properly. 6 complaints have been received in which 3 have been reported with a delay up to 10 minutes.